If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 21, Number 3, 2016

Evaluation of mechanical and geometric accuracy of two different image guidance systems in radiotherapy

Nithya Kanakavelu, Anand Mambakam Ravindran, Emmanvelrajan James Jebaseelan Samuel

Summary:

Aim

To assess the mechanical and the geometric accuracy of two different clinically used image guidance systems in radiotherapy for a period of 6 months.

Background

With the image guidance procedures being routine in the clinical radiotherapy department, the quality assurance tests for these systems become essential. The mechanical and geometric accuracy of these systems are crucial since it directly affects patient treatment set-up and delivery.

Materials and methods

We have assessed the mechanical and the geometric accuracy of two different image guidance systems (MV and kV based), being used clinically for a period of 6 months. The quality assurance tests such as imager positioning/repositioning, imaging and treatment beam isocentre coincidence, imager mechanical alignment, image scaling, geometric accuracy of cone beam computed tomography system, automatic image registration and offset calculation accuracy were assessed in this period.

Results

It was found that both systems were mechanically and geometrically accurate within ±2 mm in this period.

Conclusion

The quality assurance tests for MV based image guidance system were simple compared to kV based systems. We recommend performing periodic quality assurance tests to verify the integrity of both image guidance systems.

Signature: Rep Pract Oncol Radiother, 2016; 21(3) : 259-265