Dear Authors,
If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 22, Number 2, 2017

Evaluation of reproducibility of tumor repositioning during multiple breathing cycles for liver stereotactic body radiotherapy treatment

Ludovic Bedos, Olivier Riou, Norbert Aillères, Antoine Braccini, Jessica Molinier, Carmen Llacer Moscardo, David Azria, Pascal Fenoglietto

Summary:

Aim

To evaluate the tumor repositioning during gated volumetric modulated arc therapy (VMAT) for liver stereotactic body radiotherapy(SBRT) treatment using implanted fiducial markers and intrafraction kilovoltage (kV) images acquired during dose delivery.

Materials and methods

Since 2012, 47 liver cancer patients with implanted fiducial markers were treated using the gated VMAT technique with a Varian Truebeam STx linear accelerator. The fiducial markers were implanted inside or close to the tumor target before treatment simulation. They were defined at the maximum inhalation and exhalation phases on a 4-dimensionnal computed tomography (4DCT) acquisition. During the treatment, kV images were acquired just before the beam-on at each breathing cycle at maximum exhalation and inhalation phases to verify the fiducial markers positions. For the five first fractions of treatment in the first ten consecutive patients, a total of 2705 intrafraction kV images were retrospectively analyzed to assess the differences between expected and actual positions of the fiducial markers along the cranio-caudal (CC) direction during the exhalation phase.

Results

The mean absolute intrafractional fiducial marker deviation along the CC direction was 1.0 mm at the maximum exhalation phase. In 99%, 95% and 90% cases, the fiducial marker deviations were ≤4.5 mm, 2.8 mm and 2.2 mm, respectively.

Conclusion

Intrafraction kV images allowed us to ensure the consistency of tumor repositioning during treatment. In 99% cases, the fiducial marker deviations were ≤4.5 mm corresponding to our 5 mm treatment margin. This margin seems to be well-adapted to the gated VMAT SBRT treatment in liver disease.

Signature: Rep Pract Oncol Radiother, 2017; 22(2) : 132-140