Dear Authors,
If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 6, Number 1, 2001

Last results and lessons of the Estro European network on quality assurance in radiotherapy

Ferreira IH, Dutreix A, Bridier A, Marre D, Chavaudra J, Svensson H.

Summary:

State of the Art: The EQUAL Dosimetry audit service, set-up by the ESTRO in 1998a, is well-known through large numbers of TL-dosimeters sent to hospitals to be irradiated in clinically relevant conditions, and read at the EQUAL Measuring Laboratory, IGR, Villejuif.
All European countries have now applied to participate (404 centres out of 880) for photon and electron beams. In relation with the IAEA, this service has been extended to 27 centres of 7 countries from Eastern and Central Europe, and the Mediterranean Basin. 757 photon beams and 277 electron beams have been checked according to the "on the beam axis" procedure.

Results and outcome:

Reference beam output results demonstrating improvements with respect to the former EC Network, and good reliability of the procedure : mean ratios of measured to stated dose of 0.997 (SD 1.8%) for photon beams, and 1.003 (SD 2.1%) for electron beams.
Useful service detecting 7% of the photon beams presenting at least one check point with a deviation > 5% (2% for electron beams, but 3 times more deviations between 3 and 5%).
Re-checks and on-site visits in 8 centres reveal inaccuracies in TPS algorithms or input data and/or in local measurements (wedge factors, collimator aperture factors, PDD's, beams calibrations).
Conclusion: A number of dosimetric problems are still observed, even on the beam axis. Improvements should be introduced and checked before considering more sophisticated treatment techniques.
a The present ESTRO-EQUAL Project (MORQA) has been funded by an EU contract.

Signature: Rep Pract Oncol Radiother, 2001; 6(1) : 42-42