Dear Authors,
If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 9, Number 2, 2004

Dosimetry of beta sources for cardiovascular brachytherapy

Wojciech Bulski, Stanisław Pszona, Maria Kawczyńska

Summary:

Introduction: Over a period of the last ten years the application of ionizing radiation in the prevention of restenosis in cardiovascular diseases has become a treatment of choice.The aim of this study was the verification of procedures used for the determination of the activity and uniformity of a P-32 wire source used in brachytherapy of cardiovascular disease.
Materials and methods: Measurements of activity and uniformity were carried out for beta P-32 radioactive sources used in the Galileo afterloading system from Guidant. Measurements of activity were carried out with an ionization well chamber. Source homogeneity was measured with radiochromic films.
Results and discussion: The activity of the source determined on the basis of measurements with a well chamber agreed within 1.5% with the activity values provided by the manufacturer for each source. The results show that the activity distribution along the source length is homogeneous within ±5%. The dose rate fall-off as a function of the distance from the source axis measured with a radiochromic film is in good agreement with Monte Carlo calculations.
Conclusions: Measurements of the activity of P-32 sources with a dedicated well chamber confirm with high accuracy the results of activity measurements provided by the manufacturer. The methods of measurement of source activity and uniformity presented in this paper assure proper quality of treatment planning in cardiovascular brachytherapy.

Signature: Rep Pract Oncol Radiother, 2004; 9(2) : 33-36

FULL TXT (PDF: 0,41MB)

« back

 
INDEXED IN:

Indexed in: EMBASE®, the Excerpta Medica database, the Elsevier BIOBASE (Current Awareness in Biological Sciences) and in the Index Copernicus.