If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 9, Number 6, 2004

Experimental evaluation of ALFARD treatment planning system for 6 MV photon irradiation: a lung case study

Asghar Mesbahi, Mahmoud Allahverdi, Hossein Gheraati, Ehsan Mohammadi

Summary:

Purpose: Simple inhomogeneity correction methods available in a number of currently applied treatment planning systems are not accurate
enough for dose calculations in lung irradiations. The purpose of this study was to evaluate the accuracy of the ALFARD treatment planning system in dose calculations for lung irradiation.
Material and methods: An anatomic thorax phantom and a 6 MV photon beam were used for our irradiation. Our set-up consisted of an anterior single field for the left lung of a thorax phantom with field sizes of 5 x 5 cm and 10 x 10 cm . The percentage depth doses for each point in the lung were measured by a Pinpoint ionization chamber and calculated by the ALFARD treatment planning system. The results of calculations and measurements were compared.
Results: The ALFARD calculations overestimated measurements at all points and field sizes. The magnitude of error increased with depth
2 2 of the calculation point from 2.7% to 17.3% for the field size of 5 x 5 cm. The error for 5 x 5 cm was approximately twice as high as that
2 for 10 x 10 cm.
Conclusions: The ALFARD treatment planning system cannot calculate the dose in the lung accurately. This may be due to inherent deficiencies
of the effective path length method, which is implemented in the ALFARD treatment planning system.

Signature: Rep Pract Oncol Radiother, 2004; 9(6) : 217-221