If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 10, Number 5, 2005

A comparison of the dosimetric properties of The Electronic Portal Imaging Devices (EPIDs) LC250 and aS500

Bożena Woźniak, Małgorzata Ganowicz, Adam Bekman, Zbigniew Maniakowski

Summary:

Aim The aim of this paper was to compare the dosimetric properties of two different electronic portal imaging systems (EPID) used in the verification of doses.
Materials/Methods The Portal Vision (PV) LC250 and aS500 machines were used. The Portal Vision device is liquid-fi lled while the second machine is based on amorphous silicon. Stability and the reproducibility of signals were investigated. The relationship between the readings from EPID devices and the exit dose rate was established. Dependencies between the responses of the EPID signal and the field size, phantom thickness and source-detector distance were studied. The EPIDs measurements were compared with those from an ionization chamber in the slab phantom. The relationships were described using mathematical functions. All measurements made were based on a 6MV linear accelerator photon beam CLINAC 23EX (Varian).
Results Both devices are characterized by good stability and reproducibility. The dosimetric characteristics of EPIDs are different.
Conclusions The Portal Vision LC250 and the PV aS500 are attractive tools for dosimetry purposes. In the case of the liquid-filled PV, deformation of the beam image profile by setting-up of the gantry at angles other than 0 or 180 degree was observed.

Signature: Rep Pract Oncol Radiother, 2005; 10(5) : 249-254