Dear Authors,
If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 14, Number 4, 2009

Dosimetric verification of dose optimisation algorithm during endovascular brachytherapy of the peripheral vessels

Grzegorz Zwierzchowski, Julian Malicki, Janusz Skowronek

Summary:

Aim

Dosimetric verification of the dose optimisation model used in endovascular brachytherapy, evaluation of the optimised dose distributions using elaborated indices.

Background

The equipment used for standard radiotherapy is used in vascular brachytherapy for prevention of restenosis after angioplasty.

Material and Method

A paraffin-wax phantom, thermoluminescent detectors and MD-55 Gafchromic® films were used for dose measurements. The edge dose index (EDI), central dose index (CDI) and treatment length index (TLI) were introduced to compare dose distributions calculated and measured.

Results

Obtained values (p>0.05) show no statistically significant differences between calculated doses and measured doses. EDI values showed improvement in dose homogeneity on the edges of the application after optimisation. After optimisation CDI values from 0.9% to 1.6% for calculated and from −1.8% to 3.1% for measured showed improvement in dose homogeneity in the central part of the application. Observed values of TLI from 3% to 21% for calculated doses and from 7% to 24% for doses measured by Gafchromic® films showed increase of RIL for optimised treatment plans.

Conclusions

1/ The designed phantom allowed repeatable dosimetric verification of dose distributions in endovascular brachytherapy. 2/ Measurements with thermoluminescent detectors and Gafchromic films proved the accuracy of the calculation algorithm in endovascular brachytherapy conditions. 3/ Elaborated indices were found to be a useful tool in describing dose homogeneity. They allowed the process of optimisation to be controlled and thus an increase in dose homogeneity by 30% at the edges and by 7% in the middle of the treated volume to be achieved.

Signature: Rep Pract Oncol Radiother, 2009; 14(4) : 114-121