If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 14, Number 5, 2009

Comparison of Peripheral Dose measurements using Ionization chamber and MOSFET detector

Gopiraj Annamalai,  Ramasubramanian Velayudham

Summary:

Background

In radiation therapy, the peripheral dose (PD) – the dose outside the geometric boundaries of the radiation field – is of clinical importance. A metal oxide semiconductor field effect transistor (MOSFET) detector is used to estimate the peripheral dose.

Aim

The aim of this study is to investigate the ability of a MOSFET dosimetry system to accurately measure doses in peripheral regions of high energy X-ray beams.

Materials & Methods

The accuracy of the MOSFET system is evaluated by comparing peripheral region dose measurement with the results of standard ionization chamber measurements. Furthermore, the measurement of PD using a MOSFET detector helps us to keep the tolerance dose of any critical organ closer to the treatment field within the acceptable limits. The measurements were carried out using a 0.6 cc Farmer type ionization chamber and MOSFET 20 dosimetry system for field sizes ranging from 5 × 5 cm2 to 20 × 20 cm2 at three depths of 1.5 cm, 5 cm and 10 cm in a blue water phantom. PD were measured at distances varying from 1 cm to 30 cm from the field edges along the x axis for the open fields, with collimator rotation and with beam modifiers like 15 degree, 30 degree and 45 degree wedges.

Results

The results show a good agreement of measured dose by both methods for various field sizes, collimator rotation and wedges.

Conclusion

The MOSFET detector has a compact construction, provides instant readout, is of minimal weight and can be used on any surface.

Signature: Rep Pract Oncol Radiother, 2009; 14(5) : 176-183