If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 14, Number 6, 2009

Optimal planning parameters for simultaneous boost IMRT treatment of prostate cancer using a Beam Modulator TM

Sally J. Derbyshire, Andrew M. Morgan, Rosemary C.A. Thompson, Ann M. Henry, David I. Thwaites

Summary:

Objective

To determine the optimum energy and beam arrangement for prostate intensity-modulated radiation therapy (IMRT) delivery using an Elekta Beam Modulator™ linear accelerator, in order to inform decisions when commissioning IMRT for prostate cancer.

Methods

CMS XiO was used to create IMRT plans for a prostate patient. Arrangements with 3, 5, 7, 9 and 11 equally spaced fields, containing both a direct anterior and a direct posterior beam were used, with both 6 MV and 10 MV photons. The effects of varying the maximum number of iterations, leaf increment, number of intensity levels and minimum segment size were investigated. Treatment plans were compared using isodose distributions, conformity indices for targets and critical structures, target dose homogeneity, body dose and plan complexity.

Results

Target dose conformity and homogeneity and sparing of critical structures improved with an increasing number of beams, although any improvements were small for plans containing more than five fields. Set-ups containing a direct posterior field provided superior conformality around the rectum to anterior beam arrangements. Mean non-target dose and total number of monitor units were higher with 6 MV for all beam arrangements. The dose distribution resulting from seven 6 MV beams was considered clinically equivalent to that with five 10 MV beams.

Conclusion

Methods have been developed to plan IMRT treatments using XiO for delivery with a Beam Modulator™ that fulfil demanding dose criteria, using many different set-ups. This study suggests that 6 MV photons can produce prostate IMRT plans that are comparable to those using 10 MV. Work is ongoing to develop a complete class solution.

Signature: Rep Pract Oncol Radiother, 2009; 14(6) : 205-213