If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 17, Number 1, 2012

Comparison of dosimetric characteristics of physical and enhanced dynamic wedges

Sathiyan Saminathan, Ravikumar Manickam, Sanjay S. Supe

Summary:

Background: Wedge filters can be used as missing tissue compensators or wedge pairs to alter
the shape of isodose curves so that two beams can be angled with a small hinge angle at a
target volume without creating a hotspot.
Aim: In this study the dosimetric properties of Varian Enhanced Dynamic Wedge (EDW) and
physical wedges (PW) were analyzed and compared.
Materials and methods: Ionometric measurements of open field output factor, physical wedge
output factor, physical wedge factor and EDW factor for photon beams were carried out.
A 3D scanning water phantom was used to scan depth dose and profiles for open and PW
fields. The 2D ionization matrix was used to measure profiles of physical and EDW wedges.
The isodose curves of physical and EDW angles were obtained using a therapy verification
film.
Results and discussion: The PW output factors of photons were compared with the open field
output factors. The physical and EDW factors were compared. The difference in percentage
depth dose for open and PW fields was observed for both photon beams. The measured
isodose plots for physical and EDW were compared.
Conclusion: The wedge field output factor increases with field size and wedge angle compared
to that of the open field output factor. The number of MU to deliver a particular dose with the
EDW field is less than that of the PW field due to a change in wedge factor. The dosimetric
characteristics, like profile and isodose of EDW, closely match with that of the PW.

Signature: Rep Pract Oncol Radiother, 2012; 17(1) : 4-12