If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 19, Number 2, 2014

Technical and dosimetric aspects of the total skin electron beam technique implemented at Heidelberg University Hospital

Frank W. Hensley, Gerald Major, Carolin Edel, Henrik Hauswald, Marc Bischof

Summary:

Aim

To give a technical description and present the dosimetric proporties of the total skin electron beam technique implemented at Heidelberg University Hospital.

Background

Techniques used for total skin electron beam irradiation were developed as early as in the 1960s to 1980s and have, since then, hardly changed. However, new measurements of the established methods allow deeper insight into the dose distributions and reasons for possible deviations from uniform dose.

Materials and methods

The TSEI technique applied at Heidelberg University Hospital since 1992 consists of irradiating the patient with a superposition of two beams of low energy electrons at gantry angles of 72° and 108° while he is rotating in a standing position on a turntable at 370 cm distance from the accelerator. The energy of the electron beam is degraded to 3.9 MeV by passing through an attenuator of 6 mm of Perspex. A recent re-measurement of the dose distribution is presented using modern dosimetry tools like a linear array of ionization chambers in combination with established methods like thermoluminescent detectors and film dosimetry.

Results

The measurements show a strong dependence of dose uniformity on details of the setup like gantry angles.

Conclusions

Dose uniformity of −4/+8% to the majority of the patient's skin can be achieved, however, for the described rotational technique overdoses up to more than 20% in small regions seem unavoidable.

Signature: Rep Pract Oncol Radiother, 2014; 19(2) : 135-143